Xcell-3+ MicroElectrode Amp
Catalog Number: 40-43-1
Built in impedance check circuit with digital display and audio amplifier with a noise filter have been added to our single channel microelectrode amplifier for extracellular recording which features sharp, adjustable high and low cutoff filters, a notch (line) filter, as well as gain to x100,000. Also includes a switchable input to permit stimulation or lesion currents to the recording electrode.
Provided with metal and glass microelectrode holders which plug into the high impedance probe built into a stereotaxic holder.
The 40-43-1 Microelectrode Amplifier represents a new standard of low noise, high impedance amplification versatility for extracellular applications. State-of-the-art integrated circuits have enabled us to lower noise while increasing band width, while, at the same time, adding features that make microelectrode recording more convenient. By connecting a high impedance, low noise, differential preamplifier directly to the electrode carrier, external noise pickup is minimized. A convenient, plug-in holder for metal microelectrodes is included with the system. The probe has been designed to be used with other physiological electrodes as well. A wide range of low and high cut-off filters allows the investigator to reject signals outside the bandwidth of interest. Additionally, a sharp notch filter can be introduced to virtually eliminate interference from power lines. With a voltage gain adjustable from 10 to 100,000 in thirteen calibrated steps and a wide dynamic output range of +10V, the Xcell 3+ can be interfaced with any data collection or storage system.
AMPLIFIER:
Input: Differential
Input Impedance : 10
Input Bias Current: <4pA
Common Mode Rejection: 95dB at 60Hz
Gain: x10 to x100,000 in 13 calibrated steps (includes x10 gain from probe preamplifier)
Frequency Response: .1Hz - 15Hz (-3db)
Filter Cutoff Slopes: -12dB/Octave
High Cutoff Ranges (Hz) (-3db): 50,100, 500,1K,2K,5K,15K
Low Cutoff Ranges (Hz) (-3dB): .1, 1, 10, 50, 100, 500, 1000
Input Dynamic Range: +1V
Output Dynamic Range: +10V
Output Offset: Adjustable from -1.5 to +1.5 Volts
Cal Signal: 100Hz square wave provided equivalent to 100 mV at probe input.
Line Filter: 50/60Hz (specify) -40dB attenuation
IMPEDANCE CHECK:
Impedance: 3 digit red LED display
Measurement Ranges: 1-999 KOHm; 0.1-9.99 MOhm (1-99 MOhm optional)
Measurement Current: 10mA, peak-to-peak
Frequency: 1000Hz
AUDIO:
Noise filter: provided, on/off switch included
Sensitivity (gain before noise filter): adjustable with single turn potentiometer
Volume (gain after noise filter): adjustable with single turn potentiometer
DIMENSIONS, POWER:
Module: 7"h x 8 1/4"w x9"d (18 x 21 x23cm); 6.5 lbs. (3kg)
Probe: 1.90" x 1.2 x .62" (5 x 2.9 x 1.5cm) with 8' (2.5m) cable (other lengths optional)
Power Requirements : 115V/230V, 50-60Hz