HiZx8 8-Channel Differential Amplifier
Catalog Number: 40-41-2
Eight channel extracellular includes sharp, adjustable high and low cutoff filters, a notch (line) filter, and gain to x100,000. Features probe which can be mounted near recording site to minimize pick-up noise.
As research investigators utilize more and more microelectrodes, the need for inexpensive multi-channel amplifiers which still offer critical performance and packaging features increases as well. This lead to the development of our HiZx8 Amplifier.
The instrument features an 8-channel differential probe which can be mounted near the preparation to reduce noise artifacts. Since many experiments involve transducers which generate high frequency interference, the system includes an extra sharp, 24dB/octave, 10KHz low pass filter in additional to the adjustable high cutoff filters. A wide gain range of 10-100,000 and built in 50 or 60Hz notch filter add additional tools to combat noise issues.
Compact, with built-in power supply, two HiZx8 Amplifiers (16 channels) can be mounted in only 7 inches (17.7cm) vertical space of a standard instrument rack.
The HiZX8 provides 8 channels of high impedance microelectrode recording amplification that feature:
- A low noise x10 gain probe mounts near the preparation
- Wide gain range (10-100,000)
- Sharp (-12DB/Octave) high and low cutoff filters
- Mains line filter, (SPECIFY 50 OR 60 Hz) provided
- Built-in CAL Signal (100 Microvolts)
- Extra 24DB/Octave filter above 10kHz (Other values can be specified)
Inputs: Differential
Input Impedance: 1012 Ohms
Input Bias Current: <4pA
Common Mode Rejection: 95dB at 60Hz
Gain: x10 to x100, 000 in 13 calibrated steps (includes x10 gain from Probe preamplifier)
Frequency Response: 100Hz - 10KHz (-3dB)
Filter Cutoff Slopes: -12dB/octave plus an additional -24dB/octave above 10KHz (other value can be specified).
High Cutoff Ranges (Hz)(-3dB): 5K, 7K, 10K
Low Cutoff Ranges (Hz) (-3dB): 100,300,500
Input Dynamic Range: +1V
Output Dynamic Range: +10V
Cal Signal: 1000Hz square wave provided equivalent to 100 microvolts at Probe input
Line Filter: 50 or 60Hz (specify), -40dB attenuation
Module: 7"h x 8.25"w x 9"d (18 x 21 x 23cm). 6.5 lbs. (3kg)
Probe: 2.75"w x 4.25"l x 2"d (7 x 11 x 5cm), 2 lbs (.9 kg) with 6' (2m) cable (other lengths optional)
Power requirements: 115V/230V selectable, 50-60Hz, 1A at 115V,
40-41-2
Probe available separately.