Guideline 4000 LP+™
Catalog Number: MT-LPP
"FHC's Guideline LP+ system provides robust clinical recordings with all the flexibility of a research device."
-Jens Volkmann, MD, PhD, Chairman and Professor of Neurology, University Clinic of Würzburg, Germany
Click here for NEW LP+ 2.0 FEATURE BENEFITS
Stay on the cutting edge of cranial targeting with the feature-rich Guideline 4000 LP+. The easy-to-use, familiar interface simplifies target localization without limiting data analysis and export options. The LP+ handles the data so you can focus on the patient.
New 2.0 Software release offers ease of use:
- New Audio Equalizer for crisp, clear sound
- Simultaneous Stim / Record functionality without changing cables
- Custom stimulation design for testing new protocols
- Automatic Stop / Record at preset intervals
- Full Support, Service and Training – 24/7
Stay on the cutting edge of cranial targeting with the feature-rich Guideline 4000 LP+. The easy to use, familiar interface simplifies target localization without limiting data analysis and export options-- the LP+ manages the data while you focus on the patient.
Scaled to your Needs – The LP+ can be built with 1-5 channels for recording, stimulation, event tagging, LFP and spike analysis, and signal replay; up to 10 channels available for research users. Choose a portable case or cart with your complete system.
Quick Procedural Set-Up – Customizable preferences streamline set-up in the operating room. FHC will work with your team to pre-configure profiles for frameless, single and array configurations, as well as individual screen and report settings.
Crisp Audio – The LP+ delivers clear signals needed to navigate to target. New audio equalizer allows fine-tuning of the acoustical mix.
Stimulation Control – New features allow simultaneous stimulation and recording, and ability to test custom-designed stimulation protocols. Performing high-level research has never been easier.
Data Analysis and Export options – The LP+ has ample options for exporting data to MATLAB. Clinical users can now connect to their Waypoint Navigator to export depth and spike rate data for track visualization and analysis.
Reliable – The LP+ is durable and robust. FHC’s technical team provides on-site training and installation to ensure optimal performance and is available 24/7 to answer questions and provide operational support.
"I have worked with all of the commercially available microelectrode recording (MER) systems, and I think that the Guideline System is the best. The 2.0 software user interface is intuitive, the recording quality is excellent, and the micro- and macro- stimulation capabilities are unparalleled….”
- Michael S. Remple, PhD, Sr. Clinical Neurophysiologist, Asst Director of DBS program, Sentient, Inc.
User profiles facilitate rapid operating room setup
- Save equipment, recording, stimulation, and display preferences
- Save multiple profiles on the same Guideline 4000™ system
- Transfer profiles between systems
- Factory-specified profiles provide known, “good” configurations
- Onsite technical support sets initial profiles during system installation
A “raster” tab for viewing snapshots of all recorded events simultaneously
- Recorded snapshots are displayed according to depth
- Snapshots can be classified, color-coded, and replayed from the raster tab
- Analysis mode plots root-mean-square voltage versus depth for raster snapshots, providing an additional track selection/target identification tool
Local field potential recording capabilities (with optional hardware upgrades)
- Simultaneous low-frequency population and high-frequency spike data displays
- Fully adjustable digital filters, sampling rate, adaptive line noise filter, and gain for each channel
- Optional preamplifier hardware allows recording from 0.1 Hz-20kHz, 48 kHz sampling rate per channel
- Optional “breakout box” to independently feed out channels to downstream recording/analysis systems
- Export data directly to Matlab, Plexon or other data formats
Enhanced display configurability
- Resize, undock, and move signal analysis windows and channel thumbnails
- Drag windows and thumbnails to a second display
- Drag auxiliary channels (for accelerometer, EKG, etc.) into analysis windows to change zoom and to resize
- Switch between triggered spikes mode, chart recorder (continuous/oscilloscope) mode, and raster view mode
- Simultaneously display spikes and chart recorder data for any channel
Improved stimulation and impedance check functions
- Check impedance on multiple channels with one button press
- Stimulate up to 60 seconds with one button press
- Improved stimulation current readout
Contact FHC for free demo or on-site neuroCase trial
Applicable Standards:
General Standard: EN60601-1 General Electrical Safety
Collateral Standards: 60601-1-1, 60601-1-2, 60601-1-4, 60601-1-4, 60601-1-6
Particular Standard: EN60601-2-26
The Guideline 4000 LP+ is a Class 1 medical device with a Type BF applied part.
Physical Specifications
Dimensions:
Height: 5.5” (bottom panel to top panel)
Width: 17” (left panel to right panel)
Depth: 15.25” (front panel to back panel)
Weight: 16.85 lbs (enclosure only)
Power Requirements
100-240 VAC, 50/60 Hz, 2A
Microelectrode Recording Channels
Analog signal conditioning
Cable: 66-EL-LP or equivalent only.
Inputs: Differential or single ended Software configurable
AC/DC coupled, factory configurable
Input 1013Ω ║ 0.2pF AC, 200MΩ DC Differential
Impedance: 1013Ω ║ 7.0pF AC, 200MΩ DC Common Mode
Gain:
Preamp: 54 X
APM: 0.25X to 665X user adjustable
MER Preamp Configuration:
Upper frequency cutoff: 20kHz, 40dB/decade roll-off
Lower frequency cutoff 125Hz, 20dB/decade roll-off
Ultra-wide Preamp configuration:
Upper frequency cutoff: 20kHz, 40dB/decade roll-off
Lower frequency cutoff 0.2Hz, 20dB/decade roll-off
Auxiliary Input/Output
Auxiliary Input/Output Isolation from the patient circuit: (note: auxiliary IO jacks are earth referenced and not isolated from one another)
Dielectric Strength: >=4000VAC
Creepage Distance: >=4mm
Clearance Distance: >=2.5mm
Analog Inputs:
Quantity: two
Input Characteristics:
Coupling: factory configurable AC or DC coupled
Range: factory configurable from among following: (0-16V), (-8.0V to +8.0V), (0 to 4.0V), (-2.0V to +2.0V), (0 to 1.6V), (-800mV to +800mV), (0 to 400mV), (-200mV to +200mV), (0 to 160mV) and (-80mV to +80mV)
Bandwidth: 0 to 5.0kHz
Earth referenced, non-isolated
Analog Outputs:
Quantity: Five
Range: factory configurable: unipolar (0 to 3.0V) or bipolar (-1.5V to 1.5V) if bipolar, the output may also be AC or DC coupled
Earth referenced, non-isolated
Propagation delay: 580 µsec, filters off
Digital Inputs:
Quantity: one
Input Characteristics: TTL compatible, 1KOhm pull-down internal
VIH: 2V min
VIL: 0.8V max
VI: 0 to 5V (20mA max input current)
Earth referenced, non-isolated
Stimulation
Implementation: isolated, constant current stimulator with different ranges for microstimulation, macrostimulation and impedance checking. Stimulus waveforms are automatically routed to the appropriate contact as follows:
Macrostimulation
Cathode: macro-contact of selected cathode electrode
Anode: macro-contact of anode electrode or patient common according to anode selection
Microstimulation and Impedance Check
Cathode: microelectrode contact of selected cathode electrode
Anode: patient common
Stimulus Current Amplitude:
Compliance Voltage: ±60V
Macrostimulation
Range: 0 to 10.0 mAp-p
Display Resolution: 10 µA
Accuracy: ±5%
Adjustments: continuous adjustment using remote control amplitude knob. Resolution of 0.1mA every 7.2° degrees.
Microstimulation
Range: 0 to 100µAp-p
Display Resolution: 1 µA
Accuracy: ±5%
Adjustments: continuous adjustment using remote control amplitude knob. Resolution of 1µA every 7.2° degrees.
Stimulus Waveform (built-in square-wave):
Pulse Train Polarity: user selectable from following options
Bipolar positive going leading pulse (default)
Bipolar negative going leading pulse
Unipolar positive pulse
Unipolar negative pulse
Pulse Train Frequency: 1Hz-300Hz. Pulse Duration: 20.83 µs – 2999.95 µs note that for bipolar pulses, the duration includes both phases of the pulse.
Stimulus Duration:
Manual: stimulus while user holds down the stimulate button on the remote control
User customizable stimulation trains also supported.
Timed: user selectable from among the following options: 0.5 sec, 1 sec, 2 sec, 3 sec, 5 sec, 10 sec, 20 sec, 30 sec, 60 sec
Impedance Check
Implementation: impedance measured from microelectrode to patient common, software or hardware triggered, single channel at a time, multiple channels can be measured sequentially. Impedance result is displayed on notebook display and result is stored as an event.
Measurement Stimulus:
Waveform: sine wave
Frequency Range: 100Hz ±5% to 5kHz ±5%. 1kHz default
Amplitude: 1.0µA¬¬RMS
Duration: 1.0 to 10.0 sec. adjustable
Measurement Result
Range: auto-ranging 10M, 1M, 100k
Accuracy: ±10% of full range
Resolution: 2 digits
Remote Monitoring: Yes
Data Server: All spike sweeps (discriminated or not) and all chart recorder data as well as time stamps from digital inputs
Notebook Computer
Model: Dell Precision M6400
Processor: Intel® Core™2 Duo P8600, 2.40 GHz, 1067 MHz
Memory: 4.0 GB DDR3-1055 SDRAM
Video Card: 512 MB NVIDIA Quadro FX 2700M
Hard Drive: 80 GB Hard Drive 9.5 MM, 7200 RPMFFS
Display: 17 inch WXGA+ LCD screen
Speakers:
Model: Bose Computer MusicMonitor
Power: 100-240 V, 50-60 Hz AC
Driver: magnetically shielded, full range driver
- microTargeting™ Controller
- Guideline 4000 LP+™ Cart
- Guideline 4000 LP+™ Breakout Box
- Guideline 4000 LP+™ Extra Channel
- Guideline 4000 LP+™ Single Hard Transport Case
- Guideline 4000 LP+™ Double Soft Transport Cases
- Power Assist with Display Assembly
- microTargeting™ Handheld Remote Control
- Position Display Assembly
- Accelerometer on Wrist Band
- Onsite Case Support, Including Equipment
- Accelerometer Adapter Unit (AAU)
- Guideline 4000 LP+™ One Year Service Agreement
- Wireless Headphones
- Onsite Equipment Operation
- Guideline 4000 LP+™ Installation
- Guideline 4000 LP+™ Electrode Cable
- Guideline 4000™ Software Version 1.50
- Guideline 4000 LP+™ Line Cord (USA)
- Guideline 4000 LP+™ USB to Serial Converter Package for microTargeting™ Controller
- Guideline 4000 LP+™ Training
- QuickWave Function Generator